Continuously measure concentrations of total volatile hydrocarbons (THA) in beverage-grade CO2 gas and vaporized liquid streams pursuant to ISBT Method 10.0 criteria.
Operates on 100% H2, 40/60% H2/N2 or H2/He Fuel Mixes
Individual Flow Control Systems
3 Ranges: 0.0 ppm to 99.99 ppm, 0.0 ppm to 999.9 ppm, 0.0 ppm to 9,999 ppm
Sample introduction via a vaporized liquid CO2 source, pressurized tanks and/or individual sample cylinders
Flame Ionization Detector (FID)
Individual Flow Control Systems
Pneumatic controls
Operating Temperature: 25 °C ± 10 °C
Resolution of 0.1 ppm v/v total hydrocarbons over normal temperature ranges using zero grade H2 as fuel
Safety: flame-out indication with automatic fuel and sample line shut off
Installation: bench or 19” rack
Power: 115-230 V, 50-60 Hz