Measurement |
4 terminal Kelvin / Thompson principle eliminates errors due to lead resistance |
Display |
60,000 count + sign LCD graphics panel with backlight select display value of °C or °F |
Ranges |
7 resistance ranges |
Terminals |
4mm safety sockets |
Working temperature |
0 deg C to 40 deg C |
Storage Temperature |
-20 deg C to +50 deg C |
Safety |
Conforms to EN 61010-1:2001 50V Cat 3 |
Dimensions and weight |
358mm X 269mm X 155mm |
Mass |
<5Kg (instrument only) |
Calibration |
Digital Password protected, manual or via remote interface |
Battery |
Internal, fixed NiMh battery pack, Gas Gauge circuits to monitor battery capacity. Internal automatic FAST / TRICKLE battery charger. DC input from 9V to 36V. |
Mains supply |
External mains psu 90V – 253V, 47Hz to 63Hz with interchangeable plugs |
Range |
Current |
Resolution |
FSV |
Uncertainty |
6.0000mΩ |
10A |
100nΩ |
60mV |
0.05% Rdg +0.01%FS |
60.000mΩ |
1A |
1μΩ |
60mV |
0.05% Rdg +0.01%FS |
600.00mΩ |
100mA |
10μΩ |
60mV |
0.05% Rdg +0.01%FS |
6.0000Ω |
10mA |
100μΩ |
60mV |
0.05% Rdg +0.01%FS |
60.000Ω |
1mA |
1mΩ |
60mV |
0.05% Rdg +0.01%FS |
600.00Ω |
100μA |
10mΩ |
60mV |
0.05% Rdg +0.01%FS |
6.0000kΩ |
100μA |
100mΩ |
600mV |
0.05% Rdg +0.01%FS |